OS-4像一個(gè)自動(dòng)調(diào)溫裝置,控制臭氧發(fā)生器的開關(guān)和報(bào)警,報(bào)警基于內(nèi)部可以設(shè)定的兩個(gè)設(shè)置點(diǎn),一個(gè)用來打開臭氧發(fā)生器或報(bào)警,另一個(gè)關(guān)閉發(fā)生器。另外還有0.10ppm點(diǎn)持續(xù)監(jiān)測(cè)傳感器周圍的臭氧,該濃度對(duì)人體(健康)有影響。標(biāo)準(zhǔn)量程:0.03-10ppm,可顯示到20ppm。SPDT輸出繼電器信號(hào)的能力達(dá)5安培@250VAC或驅(qū)動(dòng)一個(gè)1/8馬力的馬達(dá)。OS-4也有數(shù)字、電壓和4-20mA輸出以驅(qū)動(dòng)馬達(dá)和控
OS-4像一個(gè)自動(dòng)調(diào)溫裝置,控制臭氧發(fā)生器的開關(guān)和報(bào)警,報(bào)警基于內(nèi)部可以設(shè)定的兩個(gè)設(shè)置點(diǎn),一個(gè)用來打開臭氧發(fā)生器或報(bào)警,另一個(gè)關(guān)閉發(fā)生器。另外還有0.10ppm點(diǎn)持續(xù)監(jiān)測(cè)傳感器周圍的臭氧,該濃度對(duì)人體(健康)有影響。標(biāo)準(zhǔn)量程:0.03-10ppm,可顯示到20ppm。SPDT輸出繼電器信號(hào)的能力達(dá)5安培@250VAC或驅(qū)動(dòng)一個(gè)1/8馬力的馬達(dá)。OS-4也有數(shù)字、電壓和4-20mA輸出以驅(qū)動(dòng)馬達(dá)和控制系統(tǒng)。OS-4不可用于酸性氣體環(huán)境或強(qiáng)VOC或鹵素氣體環(huán)境。
傳感器:加熱金屬氧化半導(dǎo)體。標(biāo)準(zhǔn)使用范圍:0.03-10ppm;可顯示到19.99ppm;
靈敏度:*響應(yīng)在低于0.03ppm的少量臭氧處;
準(zhǔn)確度:~15%(取決于應(yīng)用);
響應(yīng)時(shí)間:臭氧達(dá)到傳感器后的1分鐘內(nèi);
溫度和濕度:5~40℃,20-95% RH;
電源:12-24VDC@300MA,中心+(外殼-);5.5mm/2.5mm;
儀器尺寸:128×164×38mm;
SM-2傳感器模塊尺寸:55×37×20mm;垂直安裝中心:63mm
重量:140g
從OS-4主體到SM-4傳感器模塊,電纜可遠(yuǎn)傳長(zhǎng)達(dá)100米
人身安全報(bào)警:LED指示,繼電器激活;
失電保護(hù):失電時(shí)繼電器打開(關(guān)閉發(fā)生器);
外殼:阻燃型ABS塑料,UL認(rèn)證。繼電器:UL,CSA和CE認(rèn)證;
電氣:1:電路在12-24V/300mA或以下工作;
2:過載后保險(xiǎn)絲自動(dòng)復(fù)位,極性反向保護(hù)電極,過電壓保護(hù)電極。
Acts like a thermostat to control ozone generators and alarms based on adjustable ozone concentration set-points. Ozone ppm readout via digital meter. Simple and effective.Features: - Easy to install. Minimal adjustments.
- Sensor is in a replaceable precalibrated module. Separated by a cable, the instrument is not in the high ozone area.
- Sensor reading is temperature and relative humidity compensated.
- Replacement sensor is miniature, low cost, precalibrated and also fits our A-22.
- Operates from 12-24 VDC or from AC adapters available worldwide.
- Alarms, data loggers, and other accessories available.
- Programmable turn-on and turn-off set-points.
- Indication when .1 ppm safety standard exceeded.
Typical Applications: - Monitor ozone levels in treatment storage areas for fruits, vegetables, and other agricultural products. Monitor ozonated HVAC systems. General monitoring at .03-20 ppm.
- Control the output of ozone generators using the set-point controlled relay in the OS-4.
- Use as either ozone leak alarm or ozone failure alarm for all sizes of ozone generators.
| | OS-4 host unit normally located outside of the ozone treatment area. |
|
| |
Specifications: - Range and Accuracy: .03-20 ppm of ozone.
- Accuracy: 10-15% (Depends on application).
- Temperature and humidity: 5-40°c 20-95% RH
- Measurement Principle: HMOS (heated metal oxide semiconductor) sensor.
- Size: 127 × 64 × 42 mm (5 × 2.5 × 1.625 in).
- Weight: 200 gr (7 oz)
- Power Requirements: 12-24 VDC, 300 mA. AC adapters available.
- Output control: SPDT non-latching relay. Contacts rated at 5 amps, 250 VAC.
- .1 ppm danger alarm: Panel LED and SPDT relay 5 amp contacts.
- Connections: Terminal blocks for relay output and DC input. 2.5 mm DC power jack for AC adapter.
- Indicators: Green LED: instrument on. Yellow LED: >.1 ppm. Red LED: generator on.
- Outputs: 0-2 V, 4-20mA, RS-232 digital.
- Case: Flame-resistant ABS with mounting ears.
- Not recommended for use in the presence of strong acid gases or VOC fumes.
|